No evaporation of volatile compounds into lab air
Better composition stability of the mobile phases
Less shift of retention times
No bacterial contamination
Fit any VICI cap or VICI safety cap
The air inlet valve has a check valve which blocks vapors of the solution in a bottle, preventing the vapors from escaping into the lab. The mobile phase composition remains stable, resulting in constant retention times.
The check valve regulates the pressure when the negative pressure reaches 1.5 psi or 0.1 bar. This small negative pressure prevents evaporation of the liquid (lower vapor pressure) and still guarantees the stability of the flow rate. A negative pressure which is too high can influence the aspiration of a constant flow.
Several tests have indicated that the 0.2 µm cellulose filter will keep liquid free of bacteria and dust particles over time. If you use a filter and still have bacteria growth in your liquid, it is likely due to the presence of bacteria in the deionized water used.
Air inlet valves fit on any VICI safety cap or VICI cap.