Modular 3-channel professional system for measuring pH value / redox potential, conductivity / salinity / TDS, ion concentration and dissolved oxygen (DO). Can be equipped or later expanded with 1, 2 or 3 measuring modules of your choice.
- Large, high-resolution 7-inch colour touchscreen (capacitive), very clear and easy to operate
- Unlimited modularity and flexibility: up to 3 measuring modules can be operated in parallel, expandable at any time
- Universal communication via usb, Ethernet (LAN) and RS-232
- Full-text user guidance in 10 languages with help menu
- Unique OneClickTM operating concept (individually configurable functions/processes at the touch of a button)
- Method memory
- Maximum security and GxP support (ID protection, sophisticated user management, limit values/alarms, integrated radio-controlled clock, etc.)
- Intelligent Sensor Management (ISM)
- Connection of stirrer, automatic sample changer, PC, barcode reader, keyboard, printer (local or network)
- Internal memory for 20,000 data records (measured values, calibrations, methods)
- u-Place electrode holder with absolute vertical movement
- and much more
Features of the individual modules:
pH measuring module:
- 1 to 9-point calibration with automatic buffer recognition
- 8 pre-programmed buffer sets, 1 freely programmable buffer set
- Automatic temperature compensation with Pt1000 or NTC 30 kOhm
Conductivity measuring module:
- Automatic recognition of the calibration standard
- Specification of calculated cell constants/ input of known cell constants
- Selectable reference temperature (20 or 25°C)
- Linear, non-linear temperature compensation or USP/EP mode
pH/ion measuring module:
- All features of pH measuring module, plus:
- 2- to 9-point calibration for concentration determination
- Storage of calibration data for up to 10 ISE sensors
- Direct measurement, standard or sample addition/subtraction
Measuring module DO/BOD/BOD (dissolved O2):
- 2-point calibration, 2 predefined, 1 user-defined standard
- Automatic temperature and air pressure compensation
- Optical (InLab OptiOx) or polarographic sensors (InLab 605)
- Biological O2 requirement: pre-programmed methods according to DIN 38409 and EPA
Technical data:
- DO/BOD (dissolved O2): 0.000-50 mg/L or ppm
- 0-500% Air pressure: 500-1100 mbar
- Temperature: 0-50 °C
- For further technical data see brochure
Scope of delivery S600 kit:
Basic device, 1 x O 2 - / BOD module, 2 x empty module, pH-Electr. InLab 605 ISM, uPlace electr. arm, LabX direct software, certificate